A Modified Suction Device for Applying Precise Aspiration during Surgery: Suction Tube with Air Channel
1Department of Surgery Gülhane Military Medical Academi, Ankara, Turkey
J Clin Pract Res 2008; 30(4): 207-210
Full Text PDF

Abstract

Purpose: Aspiration of intra abdominal fluid and debris is usually needed during different stages of the operation. This is necessary in order to protect the intra abdominal space from contamination and/or to remove blood or tissue fluids that may occur during the operation. It is also essential for obtaining a clean surgical area. Aspiration of intra abdominal fluid collecting in Douglas Pouch is difficult, even with the use of lower abdominal incisions such as phannensteal. Aspiration of intra abdominal fluid collecting in the Douglas Pouch becomes more difficult when an upper abdominal incision is used because of the difficulty of access to the Douglas Pouch and suctioning the omentum and intestine with the aspirator.
Materials and Methods: In this study, a modified suction device (Suction Tube with Air Channel =STAC) has been reported which allows the surgeon easier suctioning intra abdominal fluid.
Results: STAC has an additional air channel which prevents the device from suctioning intra abdominal organs such as the intestine and omentum. We tested the efficacy of STAC in a simulated intraabdominal space such as a large transparent plastic bag with fresh sheep intestine-omentum and with 2000 cc of physiologic saline solution. While the suctioning of all the fluid was difficult (13-21 training) and time consuming (240-360 sec) when the conventional suction unit was used, all of the saline solution was easily and quickly suctioned when STAC was used (1 training and 30-60 sec, respectively).
Conclusion: In conclusion STAC provides a safe, fast and complete fluid extraction.


Cerrahide Tam Aspirasyon Sağlayan Modifiye Bir Aspirator Cihazı: Hava Kanallı Aspirator Ucu
1Department of Surgery Gülhane Military Medical Academi, Ankara, Turkey
Journal of Clinical Practice and Research 2008; 4(30): 207-210

Amaç: Ameliyatların çeşitli basamaklarında batın içi mayi ve partiküllerin aspire edilmesi gerekir. Aspirasyonun amacı batın boşluğunu kontaminasyondan korumak; kan ve doku sıvılarını uzaklaştırmaktır. Bu temiz bir cerrahi saha için gereklidir. Batın içindeki mayinin aspirasyonu Phannensteal gibi alt abdominal inzsizyonlar kullanılsa bile zordur. Bu işlem özellikle üst batın insizyonlarında douglas boşluğuna ulaşmadaki zorluklar ve omentumla barsak anslarının aspiratöre yapışmasıyla daha da zorlaşır. Çalışmada, cerraha batın içi mayii aspire etmede kolaylık sağlayan modifiye bir aspiratör ucu (Suction Tube with Air Channel=STAC) sunulmuştur.
Gereç ve Yöntemler: STAC batın içi orgqanların aspiratöre yapışmasını engelleyen ilave bir hava kanalına sahiptir.STAC'ın effektivitesi yeni kesilmiş koyun barsağı ve omentumu doldurulmuş şeffaf bir poşete 2000 cc mayi doldurularak oluşturulan modelde test ettik.
Bulgular: Klasik aspiratör kullanırken tüm mayinin aspire edilmesi zor (13-21 deneme)ve zaman alıcıydı (240-360 saniye),ancak STAC'da mayi kolayca (1 deneme) ve kısa sürede (30- 50 saniye) aspire edildi.
Sonuç: STAC güvenli, hızlı ve tam mayi aspirasyonu sağlamaktadır.